Luminex BioPlex XYP, LX200 User Manual (supplied by BioRad Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Equipamento Luminex BioPlex XYP, LX200 User Manual (supplied by BioRad. Luminex BioPlex XYP, LX200 User Manual (supplied by BioRad) User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 0
PART NUMBER
[BAAN ITEM NUMBER]
REV
BIO-RAD
LABORATORIES
2000 ALFRED NOBEL DRIVE
HERCULES, CA 94547
MATERIAL SPECIFICATION
(Non-Chemical)
10005042 C
WI.SUP.0797.02; Rev B D020248 Page 1 of 1
EXTENDED DESCRIPTION: Hardware User’s Manual Bio-Plex 200 System
ERP/PLM DESCRIPTION (30 Characters Max): USER’S MANUAL BIO-PLEX SYSTEM
SPECIFICATIONS (ALL DIMS ARE REF ONLY)
Manual is printed from the PDF file, 10005042C_BIOPLEX_HARDWARE_MANUAL.PDF.
Dimensions of the manual are 8.5 inch wide by 11.0 inch tall.
Page count = 47
Manual to be printed with black ink onto 20 pound white paper, no holes paper
The finished product should be double sided.
Care shall be taken to ensure that graphics displayed in the manual are faithfully reproduced.
The copies shall be bound with a clear plastic front cover and heavy paper back cover and spine. The
“Coverbind” system is acceptable.
The color of the paper used for the back cover and spine shall be WHITE
.
The original electronic form of the manual was created with Quark Xpress (Windows). The Quark Xpress file
is included in this document (D013638) as a zip archive named
“10005042C_BIOPLEX_HARDWARE_MANUAL QUARK FILES.ZIP”.
SPECIAL INSTRUCTIONS:
ATTRIBUTES
ESD Sensitive WEEE CDF WAREHOUSE ACCEPTED (DTS)
QUALIFIED PARTS
Eng R&D
MANUFACTURER
MANUFACTURER’S
PART NUMBER
SUPPLIER (VENDOR)
[BUSINESS PARTNER
NUMBER]
SUPPLIER PART
NUMBER
COMMENTS
RoHS
(Y/N)
Appr / Date
Bio-Rad Copy Center 10005042 NA NA N JP/09-28-
07
EQUIVALENT SUBSTITUTIONS AUTHORIZED:
RECEIVING
METHODOLOGY:
[INCLUDING DOCUMENTS AND CERTIFICATIONS NECESSARY IW ITH EACH SHIPMENT.]
PACKAGING &
SHIPPING:
NA
ORIGINATOR:
John Pocekay
R&D APPROVAL:
John Barich
APPROVAL DATE:
9-28-07
BP#:
BP-014523
No
Yes
No
Yes
Yes
No
Yes
No
Yes
No
x
x
x x
x
Vista de página 0
1 2 3 4 5 6 ... 50 51

Resumo do Conteúdo

Página 1 - 10005042 C

PART NUMBER [BAAN ITEM NUMBER] REV BIO-RAD LABORATORIES 2000 ALFRED NOBEL DRIVE HERCULES, CA 94547 MATERIAL SPECIFICATION (Non-Chemical) 10005042 C

Página 2

1.2.3 Mechanical Safety InformationCaution: During operation, this system contains exposed, moving parts. Risk of personal injuryis present. Keep han

Página 3

2.2 Description of System ComponentsThe Bio-Plex 200 system is comprised of the following components:• Array reader — combines 2 lasers, fluidics, an

Página 4 - Table of Contents

Table 1. Array Reader Front and Side Panel Features.Feature DescriptionSample arm The sample arm transports the sample from the 96-wellmicrotiter pl

Página 5

Table 2. Array Reader Rear Panel Features.Feature DescriptionCommunications The DB9-PIN connector is used to connect the array port P1 reader to the

Página 6

2.2.2 Microplate PlatformThe microplate platform (Figures 4 and 5) allows the automated processing of samples from a96-well microplate. The features

Página 7 - General Information

2.2.3 High-Throughput Fluidics (HTF) The Bio-Plex HTF (high-throughput fluidics) is designed to automate the introduction of sheathfluid into the arr

Página 8

2.2.4 Computer and MonitorThe Bio-Plex 200 system may be supplied with a computer. If so, please transfer the computer’sregistration to your company’

Página 9

2.3 Recommended Additional Equipment Not ProvidedSurge protectorWe recommend the use of a 6-outlet surge protector, with a minimum surge current of 1

Página 10 - Introduction

4. The maximum distance between the computer and the microplate platform and array readershould be 1.5 m (5 ft), the length of the communications cabl

Página 11

• Communication cable• Sheath fluid bottle• Sheath fluid waste bottle• 1 sample needle (11.6 cm/4.6 in)1. Unpack all components, being careful not to

Página 13 - Sheath filter

Fig. 12. Completed connections of computer and monitor to the Bio-Plex suspension array system.3.5 Connecting the Sheath Fluid and Waste Containers1

Página 14 - Communications port

Fig. 13. Waste and sheath fluid bottle connections.Note: To set up the 20 L sheath fluid cube container for use, remove tape from box and lift offthe

Página 15 - Table 5. HTF Features

162. Insure that the array reader, microplate platform, and HTF unit are turned off.3. Insert the Bio-Plex Manager CD-ROM into the CD drive of the com

Página 16 - Fig. 9. Bio-Plex reservoir

Fig. 15. Sample needle assembly.4. Remove the sample needle by grasping the needle and gently pushing up.5. Insert the new needle into the sample arm

Página 17 - Installation

Fig. 16. Sample needle dialog.5. Click Eject/Retract to eject the plate holder.6. Place the MCV plate on the microplate platform with the black arrow

Página 18 - 3.4 Array Reader Setup

Fig. 17. Sample needle assembly.11. By holding onto the needle height adjustment thumbscrew on the needle arm, manuallymove the needle so that it jus

Página 19

2. Click the Eject icon. Insert the MCV plate into the microplate platform. Click Retract.3. Choose Instrument from the main menu bar. Choose Addition

Página 20

Fig. 19. Device status window.6. After the pressure has stabilized, record the air and sheath pressure (Figure 19, arrows).Air pressure: ________ psi

Página 21 - 3.7 Software Installation

Fig. 20. HTF setup.9. Make the following connections to connect the HTF to the array reader:• Connect the sheath fluid line (blue fitting) to the She

Página 22

Note: You may use the additional waste line provided with the system to drain to a largerwaste container. The large waste container waste must be po

Página 24

15. Because the flow rate in step 14 is adjusted by altering the system pressure, it is necessaryto repeat steps 3 to 5 of Section 3.10 (Resetting Ins

Página 25 - 3.9 Initial System Priming

1 Make sure that the system has been set up according to the directions in the Millipore userguide.Hint: Make sure that the connector leading to the

Página 26 - 3.11 HTF Setup

3. Turn on the laboratory vacuum to maximum level (the same level that was used in the preceeding calibration procedure).4. Press on the filter plate

Página 27

Section 4Care and MaintenanceTable 6. Summary of Care and Maintenance.Daily Startup, calibrate, wash between plates, shut downWeekly Sonicate needle,

Página 28 - Fig. 20. HTF setup

Every MonthClean exterior surfaces Disconnect the instrument from AC power by turningoff the power switch on the rear of the array readerand micropla

Página 29 - Regulator

Air intake filter Replace the air intake filter every year. Disconnect the instrument from AC power by turning off the powerswitch on the rear of the

Página 30 - 3.12 Vacuum Manifold Setup

Section 5Troubleshooting5.1 Troubleshooting Guide for Bio-Plex®200 SystemProblem Possible Cause SolutionBio-Plex Manager™ Most Likely:has detected a

Página 31

Bio-Plex Manager Incorrect bead regions Compare bead regions in the assayhas detected a problem with selected in the protocol with those selected in

Página 32

Pressurizing in status bar of Leak in sheath bottle or Tighten sheath cap or replace sheathsoftware. cap bottle.Sheath and waste bottle Make sure that

Página 33 - Care and Maintenance

Bio-Plex Manager has Most likely:detected a change in sheath pressure. Sheath reservoir cap not Tighten sheath cap. Click OK.on securely Message shou

Página 34 - Every Year

Table of ContentsPageSection 1 General Information ...11.1 About This Manual ...

Página 35 - As Required

Blue power indicator light on 1. System not plugged in Turn on system, power strip.HTF is not on 2. Power strip not on Remove power cable and reconn

Página 36 - Troubleshooting

5.2 Troubleshooting Guide for Vacuum ManifoldMessage/Problem Causes SolutionNo Flow/no vacuum. Lid on plate Remove lid.All wells not wet, or Wet unus

Página 37

5.3 Technical SupportFor technical assistance with the Bio-Plex 200 system, including all hardware and software,contact your local Bio-Rad office or,

Página 38

Microplate Platform SpecificationsInput voltage range 100–240 V, ~2.25 A, 47–63 HzPhysical dimensions((WW xx DD xx HH))44 x 61 x 8 cm (17.3 x 24 x

Página 39

Section 7Warranty StatementThis warranty statement may vary outside of the continental United States. Please contact yourlocal Bio-Rad office for the

Página 40

Section 8Ordering Information — System AccessoriesCatalog # DescriptionGeneral System Accessories171-000050 HTF171-000055 Sheath Fluid, 20 L171-0020

Página 41

CatalogNumber Product Description171-002030 Protective Shield for Sample NeedlePreventative Maintenance Items171-002032 Air Intake Filter(accessed t

Página 42 - Section 6

CatalogNumber Product Description171-203032 MCV plate IV, for use with Bio-Plex Manager 5.0and 4.1 Software171-203050 Bio-Plex Reservoir, for use with

Página 43 - Computer Specifications

5. Drain the sheath fluid and waste containers.6. Rinse the waste container with 10% household bleach solution and drain.7. Remove all specimens, d

Página 44 - Warranty Statement

Equipment Return Information Label andDECONTAMINATION CERTIFICATEBefore Bio-Rad can accept this equipment, you must certify that it is NOT CONTAMINATE

Página 47 - Decontamination Information

Life ScienceGroup00-0000 0000 Sig 1106Bulletin 0000 US/EG Rev ABio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 4BIORAD Austra

Página 49

Section 1General Information1.1 About This ManualA Bio-Rad service engineer will install the Bio-Plex®200 system. However, the procedure is provided

Página 50

21.2.1 Electrical Safety InformationWarning: This instrument must be connected to an approved power source.Warning: Do not perform any maintenance or

Página 51 - Laboratories, Inc

Caution: Removal of the array reader cover is intended for trained service personnel only. Donot attempt to operate the instrument with the cover remo