Luminex xPONENT for MAGPIX Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Software Luminex xPONENT for MAGPIX. Luminex xPONENT for MAGPIX User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 126
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes

Resumo do Conteúdo

Página 1

xPONENT for MAGPIX Software User Manual IVD

Página 2

Table of ContentsChapter 1 Introduction ...1Softw

Página 3 - 2. Warranties

Initial StartupWhen you turn on the system for the first time, perform the following procedures.1. Adjust the Sample Probe Height2. Revive After Stora

Página 4

7. Verify that the reservoir is empty.8. In the Reservoir section, click well RB1.9. Verify that the plate is not warped. Warped plates can lead to in

Página 5

FIGURE 41.Sample Probe Height Adjustment1. Well D6 2. Well D1 3. Well RB1Revive After Storage RoutineAfter you have adjusted the sample probe height,

Página 6 - 5. Patent Disclaimer

FIGURE 42.Revive After Storate (Luminex) RoutineCalibration/Verification RoutineCalibration normalizes the settings for the system and ensures optimal

Página 7 - Software

8. Click Eject on the status bar. NOTE: To ensure that you get the necessary bead count, invert thecalibrator and verifier vials perpendicular to the

Página 8

3. Verify that kits have been activated and that the correct reagents (VER and Fluidicsreagents) have been added to the off-plate reagent block.4. Add

Página 9

• System Shutdown- Use this routine to perform a pre-defined system shutdown.• Alcohol Flush- Use this routine to run a single alcohol flush.• Sanitiz

Página 10 - Table of Contents

Lot Management TabUse this tab to import, export, delete, or create new calibration and verification lot information.You can also activate reagents by

Página 11

• Expiration - Displays the expiration date for the lot type selected in the Active section tothe left.• Calibrator - If a calibration reagent is sele

Página 12

FIGURE 47.Cmds and Routines TabRoutine Name - This list of pre-set routines and commands can be used for systemmaintenance. Some of these commands are

Página 13

Using Online Help ...

Página 14 - Chapter 1: Introduction

Plate Name - The name assigned to the plate.CAUTION: If using multiple plates users must be careful to ensure thatplates are used in the proper order.

Página 15 - General Guidelines

Clean - The probe aspirates 100µl of a cleaning reagent from any well or off-plate reservoir,and dispenses the reagent into the chamber. (Luminex reco

Página 16 - Bead (Microsphere) Handling

4. In the Commands section, click one or more commands to add to the routine. Thesecommands display in the plate image and in the Command Sequence lis

Página 17 - MAGPIX Technology

3. Add, delete, or change commands as necessary, then click Save As. NOTE: When you choose the CAL command, a Rinse command isperformed prior to the C

Página 18 - Interrogation of MagPlex

Exporting a Routine1. Open the Maintenance page, then open the Cmds & Routines tab.2. Click Export.3. In the Save As dialog box, navigate to the f

Página 19 - Sample Volume

Plate Location - Defines the location used to auto adjust the probe height. Select the checkbox for either the main plate, reservoir, and/or strip wel

Página 20 - Touch Screen

FIGURE 49.System Info TabThis tab contains the following:• Software• Version• Operating System• Licensing• Instrument Type• Serial Number• Firmware Ve

Página 21

• Drive Fluid• Waste FluidItems in this list relating to calibration and verification have one of the following states:• Passed - Indicates that the p

Página 22 - System Setup

FIGURE 50.System Status TabThis tab contains the following:Search By Log Type - Filters the type of log information. Select All, Maintenance, Security

Página 23 - Application Settings

Save - Click to open the Save As dialog box. This saves the file as a PDF. Pick a file nameand location to save the file, and click Save.Support Utili

Página 24 - LIS Settings

Revive After Storage Routine ...89Calibrati

Página 25

7. Verify the location where you want to store the file. To change the location, click Browse,then navigate to the new folder and click OK.8. Click Ex

Página 26 - Group Setup Tab

xPONENT for MAGPIX108

Página 27

Glossary21CFRPart1121 CFR Part 11 sets the criteria under which the Food and DrugAdministration (FDA) considers electronic records, electronic signatu

Página 28

CL2Refers to dyes embedded in the microsphere. Also see classificationchannel.classification channelA specific range of wavelengths in which light int

Página 29 - User Setup

MagPlex microspheresCarboxylated spheres with a diameter in the micrometer range. Alsocalled beads.median fluorescenceintensity (MFI)Used to denote ho

Página 30 - Global User Settings

RP1Refers to the dyes bound to the surface of the xMAP microsphere. Alsosee reporter channel.sampleThe mixture of assay components (microspheres, repo

Página 31 - Create User Account Screen

LUMINEX CORPORATION12212 Technology BoulevardAustin, Texas 78727-6115U.S.A.Voice: (512) 219-8020Fax: (512) 219-5195WMDEBergerweg 186085 AT HornThe Net

Página 32 - Batch Options Tab

xPONENT for MAGPIXxii

Página 33

Luminex® xPONENT® for MAGPIX® (xPONENT 4.1) IVD software was developed toimprove workflow and efficiency in the laboratory. Designed for ease-of-use,

Página 34 - Alert Options Tab

About This ManualThe conventions in this manual assume a basic familiarity with computers and a knowledgeof Microsoft® Windows® software. Commands are

Página 35

CAUTION: Modifying or deleting xPONENT system files may causedegradation of system performance. You can repair modified ordeleted xPONENT system files

Página 36 - Setup Email Dialog Box

NOTE: Refer to the product information sheet that accompanies your IVDassay for additional information.CAUTION: Do not use strong organic solvents wit

Página 37 - CSV Options Tab

normalizes the settings for the classification channels (CL1 and CL2) and the reporterchannel (RP1). Use the Luminex MAGPIX Calibration Kit to accompl

Página 38

Classification and Reporter FluorochromesMagPix beads in the calibration kit are used to autofocus the camera and calibrate the CL1,CL2, and RP1 chann

Página 39 - Archive Options Tab

© Luminex Corporation, 2011. All rights reserved. No part of this publication may bereproduced, transmitted, transcribed, or translated into any langu

Página 40 - Archiving a File

when the analyzer flushes the sample lines after sample acquisition and expelsapproximately 75 µL of sample back into the well.Examples• If you use a

Página 41 - Licensing Tab

xPONENT for MAGPIX8

Página 42 - Report Options Tab

System SetupIf you have a secure version of xPONENT for MAGPIX, you must have xPONENTadministrative privileges to perform System Setup tasks. If you d

Página 43

FIGURE 2.System Setup TabViewing System StatusThe System Monitor displays at the bottom of all xPONENT windows. It displays thephysical state of the L

Página 44 - Chapter 3: Using xPONENT

LIS SettingsYou must have the LIS version of the software to perform this task.To connect to the LIS, check Connect to LIS. In the Watch folder for in

Página 45 - Screen elements

FIGURE 3.New External Analysis Program Dialog BoxAdding an External Analysis Program1. If the program is on an external media source such as a CD or f

Página 46 - Right-click Menus

Editing an Analysis Program1. In the Installed Analysis Programs list, click the program you want to edit.2.Click Edit. The Edit External Analysis Pro

Página 47 - System Monitor

NOTE: The 21 CFR Part 11 package also provides full access to the SecurePackage functionality.FIGURE 4.Group Setup TabUsers are assigned to groups. Th

Página 48 - Home Page

Group Features - The Group Features list contains permission categories. When you selecta category from the list, the Features section displays the in

Página 49

• Export Processed Batch Results• Change Sample Load Volume During RunProtocol ManagementDelete ProtocolLot and Std/Ctrl Kit management• Create Std/Ct

Página 50

Standard Terms and Conditions for Use of InstrumentProductBy opening the packaging containing this product ("Product") or by using such Prod

Página 51 - System Initialization

Buttons to delete or edit settings for a selected user.User Setup TabUse this tab to create or edit a user account, remove a user from the system, or

Página 52

Defining Global User Settings1. In the Global User Settings section, type a length of time (in days) for the PasswordExpiration. The default is 180 da

Página 53 - Exporting CAL or VER Kits

Creating a New User1. Click Create New User. The Create User Account screen opens.2. Type the user ID in the User ID box. The user ID is not case-sens

Página 54 - Batches Page

FIGURE 7.Batch Options TabBatch Options. This section displays the following options:• Allow running a batch with expired reagents - Allows batches to

Página 55

• Minimum bead count for obtaining results - Type any whole number from 0 to 1000000.This is the minimum bead count that must be observed by the instr

Página 56 - Using the Batches Page

FIGURE 8.Alert Options TabAlert Options - Offers three check boxes for each alert - Dialog, Email, and Sound. Thissection displays the following even

Página 57

Setup Email - Opens the Setup Email Dialog boxFIGURE 9.Setup Email Dialog BoxEnable or disable email notification by selecting or clearing the Email A

Página 58 - Create a New Multi-Batch

4. If you choose to send an email notification for a specific event, and the email address isnot already set up, click Setup Email at the bottom of th

Página 59 - New Multibatch Tab

FIGURE 10.CSV Options TabAutomatically export results CSV file when batch is complete - Automatically exportresults CSV file when batch is complete -

Página 60

Cancel - Cancels changes.Save - Saves changes.Setting CSV Options1. In the CSV and Batch Export Options section, select the options you want to apply

Página 61 - FIGURE 22

hardware not provided by Luminex. If Product is purchased from a Luminex authorizedreseller, any warranty obligations shall be provided in writing dir

Página 62 - Save Multi-batch

• File Type - Choose from Protocols, Batches, Std/Ctrl Kits, Lots, Patient Samples, orLog. The selected information appears in the list below the Arch

Página 63 - Delete Batch

5. To exclude raw data files from the archive (suggested if you are archiving Batches) selectExclude raw data files.6. Click Archive. The command exec

Página 64 - Settings Tab

Schedule TabUse the Schedule tab to enable or change the recurrence of scheduled maintenancereminders.FIGURE 12.Schedule Tab• Select the time of day t

Página 65

FIGURE 13.Report Options Tab• Type a company name in the Company box, and any additional information in the Info:box.• Click Import Logo and then Open

Página 66 - Analytes Tab

Starting xPONENTPerform the following steps if xPONENT does not launch automatically after starting you turnon the computer.• On the PC desktop, click

Página 67

xPONENT must read this manual thoroughly before operatingthe software.1. On the System Login tab, type your user ID.2. If you are using a secure versi

Página 68 - Analysis Settings Dialog Box

FIGURE 14.Screen Elements1. Page 2. Tab 3. Right-Click MenuPageThe main elements at the top of the window are Pages. Click a page to go to that page i

Página 69 - Protocols Tab

• Copy All - Copies all data• Copy - Copies only selected data• Paste - Pastes previously copied text or data into the box• Delete - Erases text or da

Página 70 - Stds and Ctrls Tab

Pause - Pauses the system after the current command completes. Pause does not stop thesystem in the middle of running a command. You cannot run anothe

Página 71 - Stds & Ctrls Tab

FIGURE 16.Home PageDaily ActivitiesSystem Initialization - Perform a system initialization routine.Shutdown - Perform a shutdown routine.Probe and Hea

Página 72 - Managing Sample Lists

indicated on the Product label, the Product has not received approval from the United StatesFood and Drug Administration or other federal, state or lo

Página 73 - Samples Tab

Adjusting the Sample Probe HeightAdjust the sample probe height to ensure that the probe drops far enough into the well toacquire sample.NOTE: Ensure

Página 74 - Create Sample Tab

CAUTION: Correct sample probe height is critical to successful sampleacquisition and calibration. Problems with the sample probe canlead to fluid leak

Página 75

Calibrate your system at least once a week using the Calibration/Verification button on theAuto Maint tab of the Maintenance page. In addition, recali

Página 76 - Performing Analysis

FIGURE 18.Select System Initialization RoutineRunning System Initialization1. On the Home page, click System Initialization under Daily Activities.2.

Página 77 - Current Batch Tab

Deleting CAL and VER Kit Information1. Open the Maintenance page, then open the Lot Management tab.2. In the Active Reagents section, select the kit y

Página 78 - Analyze Current Batch

• Protocols - Displays when Create a New Batch from an existing Protocol is clicked.• Stds & Ctrls - Displays when Create a New Batch from an exis

Página 79

Export - Exports a batchDelete - Deletes a batch.Edit - Edits a batch.Run - Runs a batch.Using the Batches Page1. Open the Batches page.2.Click one of

Página 80

NOTE: Luminex recommends that you analyze the manufacturer’s assay kitcontrols with each batch.To create a new batch from an existing protocol:1. Read

Página 81 - Select Replay Mode

9. Type a unit of measurement in the Units box.10. Type the desired bead count for each analyte in the Count box.11.If you click Apply All, this appli

Página 82 - Analyzing a Saved Batch

FIGURE 20.New Multibatch TabThis page contains the following:Multi-Batch Name - Use this field to create a name for new multi-batches.Select Pending B

Página 83 - Filter Setup Dialog Box

contract or any other theory of law or equity arising out of, directly or indirectly, the use of theProduct or by reason of Buyer's failure to pe

Página 84 - Viewing Sample Details

6. Click Save to return to the New Multibatch tab.7. Click Run to run the multi-batch. FIGURE 21.Select Pending Batch Dialog BoxUsing xPONENT47

Página 85 - Results Page

FIGURE 22.New MultiBatch Tab with Plate LayoutxPONENT for MAGPIX48

Página 86 - Results Tab

FIGURE 23.Multibatch Report Dialog BoxSave Multi-batchAfter creating a multi-batch, you can save it to the Select Pending Batch list. When saved tothi

Página 87

Running a BatchTo run a batch, click Run Batch to open the Batches page. Select the pending batch thatyou want to run, then click Run. NOTE: If the ba

Página 88

1. Open the Batches page.2. In the Pending Batches section, click the batch you want to delete, then click Delete.The Delete Pending Batch dialog box

Página 89

FIGURE 24.Settings TabThis tab contains the following:Name and description boxes - Type the name and description in the appropriate box.Acquisition Se

Página 90

• Analysis Type - Use this list to choose from the following analysis types:• None - No analysis. Select if you have your own data post-processing pro

Página 91 - Sample Details Tab

FIGURE 25.Analytes TabThis tab contains the following:Analyte grid - A grid representing each analyte from 1 to 50. Select All selects all analytes,an

Página 92 - LIS Results Tab

FIGURE 26.Analysis Settings Dialog BoxIf you selected Quantitative on the Settings tab, the default analysis formula is 5PWeighted. To change the defa

Página 93 - Reports Tab

Apply All - Applies the information in the Units and Counts fields to all analytes.You can also specify the minimum allowable bead count per well that

Página 94

End-User License Agreement (EULA) for Luminex®SoftwareThis Luminex End-User License Agreement (“EULA”) is a legal agreement between you(either an indi

Página 95

FIGURE 27.Protocols TabThis tab contains the following:Batch Name/Description - Used to name and describe a batch.Select a Protocol - Contains the pro

Página 96 - Deleting a Protocol

FIGURE 28.Stds & Ctrls TabThis tab contains the following:Apply Std/Ctrl Kit - Opens the Select Std/Ctrl Kit dialog box. The dialog box displays t

Página 97 - Lots and Kits

• Dilution - Contains the following dilution options:• 1:2 - Halves the standard from each previous iteration.• 1:10 (Log) - Computes a value of one-t

Página 98 - Editing a Lot

FIGURE 29.Samples TabThis page contains the following tabs:• Create Sample - Displays when the Create New Samples button is clicked.• Edit Samples - D

Página 99

Protocol - Displays the protocol selected in the Samples tab. If xPONENT has an LISlicense enabled, any sample details provided by the LIS also appear

Página 100 - Maintenance

FIGURE 30.Create Sample TabThis tab contains the following:Protocol - Displays the protocol selected in the Samples tab. If xPONENT has a LIS licensee

Página 101

Save - Saves changes made to the Sample list.Close - Returns to the Samples tab.Creating a New Sample List1. Open the Samples page.2. In the Sample Li

Página 102 - Revive After Storage Routine

Current Batch TabTo view the Current Batch tab, click the Results page. The Current Batch tab also displayswhen you run a batch. Use this tab to view

Página 103

Single Step - Allows analysis of one well at a time. Select or clear this check box to turn thisfunction on or off.Scroll Arrows - Use the up, down, l

Página 104 - Daily Activities

• % Recovery - A measure of how accurately your observed results match your expectedresults following regression analysis.(Observed concentration) / (

Página 105 - Auto Maint Tab

3. GRANT OF LICENSE. Subject to the terms and conditions of this EULA, Luminex herebygrants to you a nonexclusive, nontransferable, nonassignable lice

Página 106

• Maximize - Maximizes the Progress display to fill the window.• Zoom - Enable a closer look at the analyte progress.• Save Image - Opens a Save As di

Página 107 - Lot Management Tab

Select Replay ModeTo replay a batch, open the Results page, then the Saved Batches tab. Select the batchthat you want to replay and click Replay at th

Página 108 - Cmds and Routines Tab

Analyzing a Saved BatchOpen the Results page, then open the Saved Batches tab. Select the batch name, thenclick Open. The Results, Log, and Sample Det

Página 109

FIGURE 34.Filter Setup Dialog BoxThis dialog box lets you choose the saved batch that you want to display in the CompletedBatches list, based on the o

Página 110

Import - Opens the Open dialog box to choose a batch file (.mdf) to import. Select IncludeRaw Files (LXB) to import the raw data file as well.Export -

Página 111 - Creating a New Routine

Results PageOnce data is collected in a batch, observation and analysis take place in the Results page.This page contains the following tabs:• Current

Página 112 - Editing a Routine

FIGURE 35.Results TabThis tab displays the following:• Results• Statistic• Analyte• Displayed Well• Well information (Well, Sample, Run Statistic; Ana

Página 113 - Importing a Routine

• Validate - Validates an entire selected row or cell in the Results table. Average rows orcells cannot be selected. If you haven’t selected an item o

Página 114 - Probe and Heater Tab

FIGURE 36.Settings TabName and description boxes - Type the name and description in the appropriate box.Acquisition Settings:• Volume - This is the vo

Página 115 - System Info Tab

Analysis Settings - Use this section to set the analysis type, set the number of standardsand controls, select an external analysis program, and choos

Página 116

5. TERM AND TERMINATION. Your rights under this EULA are effective until termination.You may terminate this EULA at any time by destroying the SOFTWAR

Página 117 - System Status Tab

• Fit of all Standards - The standard curve will be determined by using each individualstandard replicate when calculating the standard curve. For exa

Página 118

• Date• Message• CodeLog entries are displayed in yellow if a well was acquired with a warning. Log entries aredisplayed in red if a problem occurred

Página 119 - Support Utility Tab

LIS Results TabThis tab displays saved batches containing LIS samples.FIGURE 39.LIS Results TabFilter- Opens the Filter Setup dialog box.Clear - Click

Página 120 - Viewing the Luminex Website

Reports TabUse this tab to view, generate, and print reports. xPONENT can format your batch or multi-batch results in a variety of export formats and

Página 121

• Batch Audit• Patient ReportIf you choose Data Interpretation, a Select Analytes section appears on the right anddisplays a list of the analytes in t

Página 122 - Glossary

Select a system log report, a date from the Date list, and click Generate to generate thereport.Advanced Reports - Enables a single option, User Repor

Página 123

12. In the Count box, type the total desired bead count for each analyte. Click Apply All.13. To set a bead count and the units for a single analyte,

Página 124

Exporting a Protocol1. Open the Protocols page, then open the Protocols tab.2. Select a protocol.3. Click Export. The Save as dialog box opens.4. Sele

Página 125

3. If the protocol uses standards, type the appropriate information for each standard in theAssay Standard Information section. In each analyte column

Página 126

xPONENT for MAGPIX86